Chapter A8: Quality Assurance Process and Maintenance
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A8-6
Operator Manual © 2008 General Electric Company. All rights reserved.
QAP tests require that the system be on, be idle, and that the detector not be in power save mode for at
least 10 minutes.
IMPORTANT!: Do not use the control and display panel to change the exposure settings when
performing QAP. The exposure techniques for this calibration are automatically set up by
the calibration software. These techniques are required for proper calibration and should
not be changed by the user.
Figure A8-8 Flat-field phantom
Follow this process to perform a QAP test.
1. Press the [QAP] button at the bottom of the Worklist or Acquisition screen.
The Image Quality screen appears.
Figure A8-9 Image Quality screen
2. Select Quality Control Tests from the left side of the screen (Figure A8-10).
Figure A8-10 QAP selection from left side of screen
The Quality Assurance Program screen appears.