Chapter A8: Quality Assurance Process and Maintenance
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A8-8
Operator Manual © 2008 General Electric Company. All rights reserved.
6. Follow the instructions on screen to complete this portion of the test:
Figure A8-13 Detector Check Tests screen
a) Verify that the detector is out of its holder.
b) Press [START] and wait for the detector check tests to be completed and results to be displayed.
The Detector QC Tests screen appears automatically when the Detector Check Tests are complete.
7. Follow the instructions on screen to complete this portion of the test: