B.4 Operating Environment
Temperature 20to30ºC
Humidity 10to95%Non-condensingrelativehumidity
Pressure 50to106kPa
B.5 Storage Conditions
Temperature -25to60ºC
Humidity 5to95%Non-condensingrelativehumidity
Pressure 50to106kPa
B.6 User Control Settings
Patientcontroltemperature 35-37.5ºCin0.1ºCincrements
Aircontroltemperature 20-39ºCin0.1increments
Humidity:Servo-%relativehumidity 30-95%in5%increments
B.7 Alarms
HighAirTemp 1.5ºCoverAST(airsettemperature)
LowAirTemp 3.0ºCunderAST
BabyHot 1.0ºC*overBST(babysettemperature)
BabyCold 1.0ºC*underBST
FanFailure Failureofblowersystem
AirTemp>38ºC >38ºCforAST<37ºC
AirTemp>40ºC >40ºCairtemperatureforAST>37ºC
AirProbeFailure Compartmentairprobefailure
AirProbeDisconnect Disconnectionofcompartmentairprobe
Powerfailure Powerswitchonbutnopower
Systemfailure Non-recoverablesystemfailure
AddWater Humidierwaterlevellow
Scale:WeightExceedsMaximum Greaterthan8kgload(visualonly)
Scale:ScaleFailure Detectablesystemfailure(visualonly)
*Canbere-setontheServicescreento0.5ºC
170 6600-0356-000 103 ©2001byDatex-Ohmeda,Inc..Allrightsreserved.
AppendixB:Specications

Table of Contents

Other manuals for GE Giraffe

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe and is the answer not in the manual?

GE Giraffe Specifications

General IconGeneral
ManufacturerGE Healthcare
TypeIncubator
Humidity Range30% to 95% RH
Power Supply50/60 Hz
ModelGiraffe
Featuresintegrated scale
Control ModesAir and Skin

Related product manuals