EasyManua.ls Logo

GE Giraffe - Operating Environment; Storage Conditions; User Control Settings; Alarms

GE Giraffe
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B.4 Operating Environment
Temperature 20to30ºC
Humidity 10to95%Non-condensingrelativehumidity
Pressure 50to106kPa
B.5 Storage Conditions
Temperature -25to60ºC
Humidity 5to95%Non-condensingrelativehumidity
Pressure 50to106kPa
B.6 User Control Settings
Patientcontroltemperature 35-37.5ºCin0.1ºCincrements
Aircontroltemperature 20-39ºCin0.1increments
Humidity:Servo-%relativehumidity 30-95%in5%increments
B.7 Alarms
HighAirTemp 1.5ºCoverAST(airsettemperature)
LowAirTemp 3.0ºCunderAST
BabyHot 1.0ºC*overBST(babysettemperature)
BabyCold 1.0ºC*underBST
FanFailure Failureofblowersystem
AirTemp>38ºC >38ºCforAST<37ºC
AirTemp>40ºC >40ºCairtemperatureforAST>37ºC
AirProbeFailure Compartmentairprobefailure
AirProbeDisconnect Disconnectionofcompartmentairprobe
Powerfailure Powerswitchonbutnopower
Systemfailure Non-recoverablesystemfailure
AddWater Humidierwaterlevellow
Scale:WeightExceedsMaximum Greaterthan8kgload(visualonly)
Scale:ScaleFailure Detectablesystemfailure(visualonly)
*Canbere-setontheServicescreento0.5ºC
170 6600-0356-000 103 ©2001byDatex-Ohmeda,Inc..Allrightsreserved.
AppendixB:Specications

Table of Contents

Other manuals for GE Giraffe

Related product manuals