2.5.2.2 O Center Weight Check
1. Enter the service screen by holding the override key (>37) while powering up. Set the scale resolution to 5
grams (if option is available).
2. Shut o the unit and power it back up to enter normal weighing mode.
3. Place a known test weight 10 cm (4”) from the center of the mattress in 4 positions 90 degrees from each
other (for example- toward each corner of the mattress) and check the reading at each position. Resulting
weight readings should be within +/- 10 g of previously obtained center weight reading.
2.6 Other Accessory Checks
1. Check that all accessories are securely mounted and out of the path of the hood in the open position.
2. Check the operation of any accessories with reference to their appropriate operation manuals.
3. Setup any required suction or gas supply systems. Check them for leaks as described in their respective operation
manuals.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0356-000 103 15
Chapter 2: Service Checkout

Table of Contents

Other manuals for GE Giraffe

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe and is the answer not in the manual?

GE Giraffe Specifications

General IconGeneral
ManufacturerGE Healthcare
TypeIncubator
Humidity Range30% to 95% RH
Power Supply50/60 Hz
ModelGiraffe
Featuresintegrated scale
Control ModesAir and Skin