5.5.7 Humidier Assembly Repairs
A design change was made to the humidier assembly to make it easier to remove the screws during
disassembly. Units shipped since about March 2002 have this revised design. The earlier design can be
identied by the button head cap screws used in the assembly. The later design uses Phillips head screws to
make disassembly easier.
When working on the earlier design if you have diculty with disassembly you may wish to replace the
entire humidier assembly with the later design 6600-0245-850 . When disassembling the older design, we
recommend that you replace the old screws with the new screws listed below. You may also require the other
parts listed below if they are damaged during disassembly. Refer to the Tips section 5.5.7.1.
The reservoir switch button was also redesigned to prevent it from jamming in the closed position. You should
also replace this button when working on an old style humidier.
Qty. per assy. Description Part Number
3 Screw, M3 X 8, Phillips head, Teon coated 6600-1255-401
4 Screw, M3 X 10, Phillips head, Teon coated 6600-1255-402
4 Screw, M3 X 12, Phillips head 6600-1255-403
2 Screw, M3 X 16, Phillips head 6600-1255-404
4 Spacer, .125 ID 6600-1779-500
1 Ramp block 6600-1777-500
1 Heater mount 6600-1291-500
1 Button, reservoir switch 6600-1298-500
5.5.7.1 Tips for Repairing the Humidier Assembly on an Old Style Unit
Some of the hex head screws may be dicult to remove. To minimize the chances of stripping the heads, be
sure that the 1.5 mm Allen key you are using is not worn, and is not a ball head style. If you strip the heads,
you may need to drill them out. If you drill out the heads, you may need to replace the ramp block and/or the
heater mount if you have diculty in removing the body of the screw.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0356-000 103 81
Chapter 5: Repair Procedures