Chapter 4: Troubleshooting
4.1 Startup Sequence
By default, the Girae Incubator starts up in user mode. The next section describes how to start up the
Incubator in service mode. The following table describes the Girae Incubator startup sequence for each mode:
User Mode Service Mode
1. The alarm light turns on and remains on.
2. All LEDs and segments light up and the EL
screen is yellow.
3. All LEDs turn o. The EL screen displays the
Girae logo.
4. The alarm light turns o. Software revisions
appear on the EL screen with OKs.
5. The user screen appears.
1. The alarm light turns on and remains on.
2. All LEDs and segments light up and the EL screen is
yellow.
3. All LEDs turn o. The EL screen displays the Girae
logo.
4. The alarm light turns o. Software revisions appear
on the EL screen without OKs.
5. The service screen appears.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0356-000 103 27