Item Description
Comfort Allows you to remove the Comfort Screen feature.
Set Time
Set real time clock for time, day and date. Choose how date is displayed (North American or
European). Choose a 12 hour (AM/PM) or 24 hour time display.
View Mods
Display the current software revision of the options installed on this specic unit (Humidier,
Scale, SPO2, etc.)
Down Go to second service screen.
Figure 4-2 Second Service Screen
Up Return to previous screen.
Status
Check status of all self test the
software runs continuously.
These include: incubator heater
on (INCHTRON), incubator
heater o (I/HTROFF), humidier
heater on (RHHTR), remote
monitoring data stream
(RS232LOOP), incubator fan on
(FANON), and incubator fan o
(FANOFF).
If the RS232 option is not
installed, RS232LOOP will display
fail. To test the circuit if the
option is installed, short pins 2
and 3 on the 9 pin connector
on the back of the electrical
enclosure.
Figure 4-3 Status Menu
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0356-000 103 29