Chapter 3: Calibration and Maintenance
SENSITIVE TO ELECTROSTATIC DISCHARGE CAUTION
An Electrostatic Discharge (ESD) Susceptibility symbol is displayed to alert service personnel that the
part(s) are sensitive to electrostatic discharge and that static control procedures must be used to
prevent damage to the equipment.
WARNING:
After performing any repair or calibration, always perform the Service Checkout Procedure before
putting the unit back into service.
3.1 Service Maintenance Schedule
The unit should be maintained in accordance with the procedures detailed in this manual. Service maintenance
must be performed by a technically competent individual.
This schedule lists the minimum frequencies. Always follow hospital and local regulations for required
frequencies.
Frequency Maintenance
Annually
• Perform the electrical safety and calibration procedure as described later in this
chapter.
• Perform checkout procedures in Chapter 2. If the scales fail the weight check, perform
the Scale Calibration procedure in section 3.7.
Every Two Years
• Replace the battery.
NOTE: The battery is used to sound the power failure alarm and to power memory circuits
during a power failure. Run the unit for at least two hours to charge the new battery
before using the unit. The battery charges only when the unit is powered on.
Every Three
Years
Calibratethehumidier.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0356-000 103 19

Table of Contents

Other manuals for GE Giraffe

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe and is the answer not in the manual?

GE Giraffe Specifications

General IconGeneral
ManufacturerGE Healthcare
TypeIncubator
Humidity Range30% to 95% RH
Power Supply50/60 Hz
ModelGiraffe
Featuresintegrated scale
Control ModesAir and Skin

Related product manuals