IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 3 Installation 3-57
3-9-5 Setting Up for Communication with a Prosolv Workstation (cont’d)
8) Enter the IP address for the DICOM server in the IP-address field.
9) Trackball to the Dicom Storage Server line in the Services table. The Services tab is expanded,
as shown below:
Figure 3-41 Connectivity Tab (Services for DICOM Server Expanded)
10) Enter the AE Title and the Port No of the DICOM server in the appropriate fields.
NOTE: Ensure that the Allow Raw Data checkbox is deselected.
11.)Trackball to the Dicom Print line in the Services table. The Services tab is expanded, as shown in
Figure 3-42 below.