EasyManua.ls Logo

GE VIVID 3 BT00 - VCR;ECG Tab

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 3 Installation 3-45
3-8-9 VCR/ECG Tab (cont’d)
NOTE: To test the VCR option, insert a video cassette. Refer to 4-3-2 "Peripherals" on page 4-4.
5) In the ECG and Phono Options area, complete the following, as required:
Select each connected option including, ECG Exist, ECG External, Phono Exist and/or
Phono Filter.
•In the No. of Periods to Store (auto-locate) field, enter the number of heart cycles (1-4) of the
frozen image that will be displayed in the multiple screen/split screen when in freeze mode.
In the Standard Systole Time (msec) field, enter the default systole time, which enables
cineloop synchronization.
In the Store time without ECG (msec) field, define the default length of a cineloop when there
is no ECG trace.
Select the Analyze only Systole checkbox to place markers only on the systole portion, for
synchronization and analysis purposes.
Select the Advanced Synchronization checkbox to enable the use of a special
synchronization algorithm. This algorithm allows non-linear stretching of the time base to
compensate for non-linearity of the heart cycle between the resting HR and elevated HR.
Select the Special Systole Colorization checkbox to enable the function of systole and
diastole colorization, if required. This function will present the systolic ECG signal in a different
color from that of the diastolic ECG signal.
NOTE: The default factory setups are seen in Figure 3-27 "VCR/ECG Tab" on page 3-44. However, the
user may want to set different values and presets.