EasyManua.ls Logo

GE VIVID 3 BT00 - Front End Crate Block Diagram - RFT

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 5 Components and Function (Theory) 5-9
5-4-3 Front End Crate Block Diagram - RFT
A block diagram of the Front End Crate is shown in Figure 5-1 on page 5-4.
For a detailed description of the Front End Crate components, refer to Front End DC Power Distribution
on page 5 - 44.
Figure 5-5 Front End Crate Block Diagram - RFT Configuration
The Front End can be divided into two subsystems, as follows:
1.) The Front End subsystem which includes:
- Front Board:
- MUX Board:
- BF64 (Beamformer Board)
- FEC (Front End Controller Board)
2.) The Mid Processors subsystem which includes:
- RFT (Radio Frequency & Tissue Board)
- IMP2 (Image Port 2).
Front
Board
TR4 Boards
MUX
Board
64Ch
Beam Former
A/D
Board
RFT
Board
Image Port
Board
FEC
(Front End Controller)
Board
Power Supply
Tx P.S.
L.V P.S
Vingmed manufactured boards
I C bus
2
PCI2IP
I
2
C bus
FE bus
Pipeline
link
VME bus
Front
Board
TR4 Boards
MUX
Board
64Ch
Beam Former
A/D
Board
RFT
Board
Image Port
Board
FEC
(Front End Controller)
Board
Power Supply
Tx P.S.
L.V P.S
Vingmed manufactured boards
I C bus
2
PCI2IP
I
2
C bus
FE bus
Pipeline
link
VME bus

Table of Contents

Related product manuals