EasyManua.ls Logo

GE VIVID 3 BT00 - Rotary Knob (External Keyboard) Replacement Procedure

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-41
8-3-7 Rotary Knob (External Keyboard) Replacement Procedure
NOTE: This section describes the replacement procedure for a parameter adjustment Rotary Knob on the
external keyboard.
8-3-7-1 Tools
Use a small flat screw driver, as indicated in the rotary knob replacement procedure.
8-3-7-2 Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
8-3-7-3 Rotary Knob Removal Procedure
1) Carefully place a small flat screw driver between the rotary knob you wish to remove and the
surface of the keyboard - see Figure 8-39 "Keypad Replacement" on page 8-39.
2) Use the screw driver to gently lever off the defective rotary knob, until it is completely loosened from
its base.
3.) Lift off and remove the defective knob - see Figure 8-41 below.
4.) Carefully take out the metal tension plate from the centre of the knob (refer to Figure 8-42 "Tension
Plate Fitted in Centre of Knob" on page 8-42), before discarding the defective knob.
Do not discard
the tension plate.
Figure 8-41 Rotary Knob Replacement

Table of Contents

Related product manuals