EasyManua.ls Logo

GE VIVID 3 BT00 - Electrical Requirements

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 2 Pre-Installation 2-3
2-2-5 Electrical Requirements
NOTE: GE Medical Systems requires a dedicated power and Ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE: Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-5-1 Vivid 3N Pro/Expert Power Requirements
Electrical specifications for the Vivid 3N Pro/Expert ultrasound scanner, monitor and onboard
peripherals are as follows:
2-2-5-2 Inrush Current
Inrush current is not a factor for consideration, due to the inrush current limiting properties of the power
supplies.
Maximum power requirement = 1.2 KVa
100V AC: 8A
120V AC: 8A
220 - 240V AC: 4A
2-2-5-3 Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
Table 2-10 Electrical Requirements, All Models
Voltage Tolerances Op. Current Frequency
120 VAC ±10% 8A 50-60 Hz
230 VAC ±10% 4A 50-60 Hz
CAUTION
POWER OUTAGE MAY OCCUR.
The Vivid 3N Pro/Expert requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.

Table of Contents

Related product manuals