GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-121
8-8-2 Sony VCR Replacement Procedure
NOTE: This section describes the installation and removal procedure of the SONY SVO-9500MDP2 and
SVO-9500MD2.
The SVO-9500MDP2 (PAL) and SVO-9500MD2 (NTSC) are alike, except for the Color Signal system.
The installation procedure is the same for both models.
Figure 8-104 SONY Video Cassette Recorder (VCR)
8-8-2-1 Tools
No special tools needed.
8-8-2-2 Manpower
One person - 30 mins.
8-8-2-3 Preparations
1) Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation.
2.) Remove the Connectors Panels Cover as described in 8-2-5-3 "Connector Panels Cover Removal
Procedure" on page 8-8.
3.) Unpack the VCR.
4.) Check that the VCR AC voltage corresponds to that of your local voltage.
5.) Verify that the switches on the VCR are as shown in Figure 8-105 "VCR Switches Setting" on page
8-122. If necessary, adjust accordingly.