IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 8 Replacement Procedures 8-69
8-5-5 Plug & Scan Battery Replacement Procedure
If the Plug & Scan battery is replaced, ensure that you properly transport or dispose of it according to
the procedures established by your facility, company, or regional regulatory agency.
8-5-5-1 Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrenches,
as indicated in the Plug & Scan battery replacement procedure.
8-5-5-2 Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in Chapter 3 - Installation. Voltage may
exist in the system when a Plug & Scan is present. Ensure that a full shutdown of the system is
performed, otherwise voltage might be present that can cause injury and/or damage.
8-5-5-3 Plug & Scan Battery Removal Procedure
1) Remove the left cover, as described in 8-2-2-3 "Side Covers Removal Procedure" on page 8-4.
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP cover as described 8-5-1 "Back End Processor Replacement Procedure" on page
8-57. The Plug & Scan battery and its cover are located to the right of the BEP.
4) Remove the screws that secure the battery bracket in position (see Figure 8-63 below - BEP1 has
three screws; BEP2 has two); remove the bracket.
Figure 8-63 Plug & Scan Battery Bracket and Screws
5) Disconnect the battery cable from the Plug & Scan board.
BEP2BEP1
Battery Bracket
Battery Bracket
(third screw not visible)
Screw
Screw
Screw