GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
3-44 Section 3-8 - System Configuration
3-8-9 VCR/ECG Tab
1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
2) Trackball to the VCR/ECG tab and press Select. The VCR/ECG tab is displayed, as shown below:
Figure 3-27 VCR/ECG Tab
3) In the VCR Options area, complete the following, as required:
• Select the video system from the Video Mode dropdown list, for example, PAL or NTSC.
• Select the SmartVCR option to enable Measurement and Analysis measurements and
calculations on recorded VCR images.
• Select the Patient Name on second line option to display the patient’s name on the second
line. For use when an external monitor omits picture borders.
4) If done, trackball to the OK button and press Select.