GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 3 Installation 3-61
3-9-5-2 Validating Communication with the Workstation (cont’d)
5) Press <Ctrl+S> to view the status of the spooler, as shown below:
Figure 3-44 Job Spooler
The Status column can contain the following options:
Hold: Images are in the spooler buffer but transmission has not begun as the exam is
active.
Active: Images are being transmitted over the network.
Done: Transmission is complete and all the images have been transmitted over the
network.
Fail: Transmission failed or is incomplete. An error message will be displayed in the Error
column.