GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7-33
7-3-8-3 FLA Jumper Switch (cont’d)
Figure 7-18 Report Dialog Box - FLA Jumper Test (Closed)
The Report dialog box contains the following information:
• Test Result Message: Displays the test result.
• FB Elements: Displays a graphic representation of the front board elements, as follows:
• Each element is represented by a button (highlighted in green) that displays the element,
and the channel number that is associated with that element.
• Each button is highlighted yellow when selected.
• Each button is highlighted red if errors were detected for one or more elements.
• Checkboxes: Enable you to manipulate bad channels:
• Show Only Bad Channels: Select this checkbox to display only bad channels in the
Display area.
• Mark bad channel: Select this checkbox to manually mark the currently selected channel
as bad.