IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-84 Section 7-4 - Performing Back End Diagnostics on the System
7-4-7-1 Benchmark Testing (cont’d)
Figure 7-55 Benchmark Testing Results
In the information displayed under CPU Description, Memory and Frame Rate, make sure that the
values displayed in the upper line (which indicates current computer results) are higher or equal to those
indicated in the lower line.