IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-116 Section 7-5 - Common Service Interface
7-5-8-2 Scanner Utilities
The functions accessed under Scanner Utilities are described below.
System Shutdown - There are three options provided
(described on previous page - also see Figure 7-88)
Reset Database - used to empty the database and reset a new one
(see Figure 7-89 below).
Figure 7-89 Reset Database
WARNING: THIS PROCESS IS IRREVERSIBLE! Before proceeding, make sure you have a
complete backup of the existing database.
When selecting Reset Database, the following warning message is displayed:
Click OK to proceed.