GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
7-126 Section 7-5 - Common Service Interface
7-5-8-4 iLinq IP Save/Restore
The iLinq IP Save/Restore utility, which is accessed from the System Configuration window (under the
Technical Support Tab), will allow you to save the iLinq IP settings to the hard disk, as described below.
7-5-8-4-1 iLinq IP Save Procedure:
1.) In the System Configuration window, click on the Technical Support tab - see Figure 7-99 below.
2.) Trackball to iLing Settings: Save and press Select. The system saves the presets to the disk.
3.) T
RACKBALL to OK and press Select. The scanning screen is re-displayed.
7-5-8-4-2 iLinq IP Restore Procedure:
When required (for example, after software re-installation), you can perform Restore of iLinq IP settings
from the hard disk, as follows:
1.) From the System Configuration window (under the Technical Support Tab - refer to Figure 7-99,
above) trackball to iLinq Settings: Restore and press Select.
2.) A blank screen appears; wait for a while until the phone and modem options message is displayed
then T
RACKBALL to OK and press Select.
3.) Wait while the restore process is running - this may take a while. When complete, a message is
displayed prompting for system shutdown.
4.) Perform a full system shutdown, then re-start the Vivid™ 3 system. The restored iLinq IP settings
will now be operational.
Figure 7-99 Technical Support Tab