IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-18 Section 8-2 - Cover Replacement Procedures
8-2-10-3 Control Console Bottom Cover (Upper Section) Removal Procedure (cont’d)
7) Loosen the two screws fastening the AC/VGA connectors box. These screws can be accessed from
the opened space at the rear of the gas spring using a long screw driver, as shown below:
Figure 8-19 Removing Screws from the AC/VGA Connectors Box
8) Remove the four screws fastening the metal cover under the front part of the control console, and
remove this cover.
Figure 8-20 Removing the Metal Cover and Bottom Cover Screws
9) Remove the six screws fastening the bottom cover, three on either side.
10) Remove the bottom cover towards the rear of the machine.
Metal
Cover
Screws
Bottom
Cover
Screws
(one side)