GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
8-100 Section 8-6 - Lower Section Components Replacement
8-6-6-3 Up/Down Handle Removal - Procedure 1 (cont’d)
3.) Extract the M6 screw from the up/down handle by sliding it out.
4.) Release the gas spring cable from the broken up/down handle - see Figure 8-83 "Gas Spring
Release Handle" on page 8-100.
5.) Remove the broken up/down handle.
Figure 8-83 Gas Spring Release Handle
Pic to be Replaced
Gas Spring Cable