EasyManua.ls Logo

GE VIVID 3 BT00 - Page 580

GE VIVID 3 BT00
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
10-16 Section 10-6 - Probe Maintenance
10-6-4-3 Cleaning and Sterilizing Probes Used Externally
10-6-4-4 Cleaning and Sterilizing Probes Used Internally
Table 10-12 Cleaning & Sterilizing Procedure for Probes Used Externally
Step Description
1 To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
2 Do not immerse the connector or its strain relief in solution. Use ultrasound probe cleaning towels.
Wash the probe and probe cable in warm soap and water solution (below 80°F). Use only mild
non-abrasive soap.
3 Follow the manufacturer’s cold sterilization instructions to disinfect the probe, including concentration
and soaking time; 10 minutes is a typical disinfecting period.
4 Follow step 3 for a much longer duration to sterilize the probe; 10 hours is a typical cold chemical
sterilization duration.
5 When complete, thoroughly rinse the washed lens in clear water. Apply sterile distilled water for at
least 60 seconds. Air dry.
6 Check the lens for cracking and the cable insulation for separation.
7 Store the probe in its carrying case.
Table 10-13 Cleaning & Sterilizing Procedure for Probes (TEE) Used Internally
Step Description
1 To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
2 Do not immerse the connector or its strain relief in solution. Use ultrasound probe cleaning towels.
Wash the probe and probe cable in warm soap and water solution (below 80°F). Use only mild
non-abrasive soap.
3 Follow the manufacturer’s cold sterilization instructions to disinfect the probe, including concentration
and soaking time; 10 minutes is a typical disinfecting period.
4 Follow step 3 for a much longer duration to sterilize the probe; 10 hours is a typical cold chemical
sterilization duration.
5 When complete, thoroughly rinse the washed lens in clear water. Apply sterile distilled water for at
least 60 seconds. Air dry.
6 Check the lens for cracking and the cable insulation for separation.
7 Store the probe in its carrying case.

Table of Contents

Related product manuals