IRECTION FK091075, REVISION 04 VIVID 3N PRO/EXPERT SERVICE MANUAL
10-32 Section 10-9 - PM and Safety Inspection Certificates
Vivid 3N Electrical Safety Inspection Certificate
Electrical Test Performed
Max Value
Allowed
Value
Measured
OK? Comments
Outlet (correct Ground & wiring config.)
System Ground Continuity
Chassis Source Current Leakage - Probe
Chassis Source Current Leakage - Caster
Chassis Source Current Leakage - CRT
Patient Lead Source Leakage
(Lead-to-Ground)
Patient Lead Source Leakage
(Lead-to-Lead)
Patient Lead Source Leakage
(Isolation)
Peripheral 1 Current Leakage
Peripheral 1 Ground Continuity
Peripheral 2 Current Leakage
Peripheral 2 Ground Continuity
Peripheral 3 Current Leakage
Peripheral 3 Ground Continuity
PROBES
Probe Number
(from page 10-31)
Max Value
Allowed
Max Value
Measured
OK? Comments
Probe 1:
Probe 2:
Probe 3:
Probe 4:
Probe 5:
Probe 6:
Probe 7:
Probe 8:
Probe 9:
Final Check: All system covers are in place. The Vivid 3N system scans with all probes, as expected.
Accepted by: