Chapter 4 - General procedures and Functional checks 4 - 7
4-2-1-3 Turn Unit ON (cont’d)
i.) As soon as the software has been loaded, either a 2D screen is displayed on the screen,
indicating that a probe has been connected, or a No Mode screen is displayed, indicating that
no probe has been connected.
NOTE: Total time used for start-up is typical one and a half minutes or less. If starting after a power loss
or a lock-up, the start-up time may be up to four minutes.
Figure 4-4 2D Screen on the display.