7 - 2 Section 7-2 - Service Safety Considerations
Section 7-2
Service Safety Considerations
Section 7-3
FAQ - Frequently Asked Questions
7-3-1 High System Temperature Error
Question:
What is the highest temperature when VIVID E9 starts giving the “High System Temperature” Error?
Answer:
• All Card Rack (PCBs, or Front End) sensors:
- The limit for the first warning is 78 degrees centigrade (78 ºC).
- The system starts the shutdown sequence at 80 degrees centigrade (80 ºC)
• New in v110.1.10 - DRX_TOP sensors:
- The limit for the first warning is 87 degrees centigrade (87 ºC).
- The system starts the shutdown sequence at 90 degrees centigrade (90 ºC).
• New in v112.0:
- There is a new reminder showing “Clean air filter”.
It is displayed at start-up of the VIVID E9, and is repeated at specific pre-defined intervals.
7-3-2 Reset the BEP from a Hang
Question:
It seems that the BEP is “hanging” - it’s not responding at all. How do I Reset it?
Answer:
1.) Try this method first:
Press the ON/OFF switch on the Operator Panel for more than six seconds. This should cause the
VIVID E9 to perform a “Forced Restart”.
2.) If the instruction in step 1 didn’t help, as the last solution, after waiting several minutes, switch off
the power on the rear of the system.
DANGER
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
WARNINGWARNING
IF THE COVERS ARE REMOVED FROM AN OPERATING VIVID E9, SOME METAL
SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF
TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
WARNINGWARNING
USE ALL PERSONAL PROTECTION EQUIPMENT (PPE) SUCH AS GLOVES, SAFETY
SHOES, SAFETY GLASSES, AND KNEELING PAD, TO REDUCE THE RISK OF INJURY.