GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 4 - General procedures and Functional checks 4 - 17
4-2-6 Moving and Transporting the VIVID E9
4-2-6-1 The Casters (Wheels) control
The wheels of the unit are controlled by the pedals located between the front wheels of the unit.
Examine the wheels frequently for defects to avoid breaking or jamming.
1.) Press the right pedal to engage the Parking brake.
2.) Press the center pedal to release the Parking brake.
3.) Press the left pedal to engage the Swivel lock.
4.) Press the center pedal to release the Swivel lock.
4-2-6-2 To prepare the unit to be moved
1.) If not locked, move the keyboard console and LCD monitor to the park position (see: 4-2-3 "Top
Console position adjustment" on page 4-11 and 4-2-4 "LCD Monitor position adjustment" on page
4-14).
2.) Turn the system off, including the circuit breaker (see 4-2-2 "Power shut down" on page 4-8), and
remove the plug from the wall.
3.) Disconnect all cables linking the unit to any off-board peripheral devices and network.
4.) Secure the unit’s power cable.
5.) Place all probes in the probe holder. Ensure that the probe cables do not protrude from the unit or
interfere with the wheels.
6.) Ensure that no loose items are left on the unit.
7.) Fold down the monitor.
8.) Unlock the brake.