Chapter 3 - System setup 3 - 5
3-3-4 Receiving the VIVID E9
3-3-4-1 Examine all packages
Examine all packages closely at time of delivery, as described in the procedure below.
Table 3-3 Examine all packages
STEP TASK ILLUSTRATIONS
1
Is damage apparent?
• If yes; continue with the instructions in subsection
3-3-4-2 - Damage in transportation.
• If no; continue with the next step.
2 Is the Shock Indicator red colored inside the middle of the
indicator?
• If yes: The Shock Indicator has been activated.
Continue with the instructions in subsection
3-3-4-2 - Damage in transportation, then continue with
the next step.
• If no: continue with the next step.
3 Is the Tilt Indicator red colored inside the middle of the
indicator?
• If yes: The Tilt Indicator has been activated.
Continue with the instructions in subsection
3-3-4-2 - Damage in transportation before you continue
with the next step.
• If no: continue with the next step
4 Wooden box ONLY:
The two upper hinges on the Front Side and the Rear Side of
the wooden transportation box have been sealed with red
plastic seals, marked GE Vingmed Ultrasound and a serial
number. Verify that the four red plastic seals are intact at
arrival.
If seals are broken:
If broken, it indicates that the container may have been
opened after it left the manufacturer.
Continue with the instructions in subsection
3-3-4-2 - Damage in transportation.
Continue with the instructions in 3-3-5 - Unpacking VIVID E9
from the Wooden Box.
5 Carton Box ONLY: Continue with 3-3-6 "Unpacking
VIVID E9 from the Carton Box" on page 3-12.
RED COLOR