7 - 20 Section 7-9 - Troubleshooting
7-9-9 System Temperature Too High
When the temperature inside the VIVID E9 increase, the fan speed will increase to cool down the
system. If the air filters become too dusty, a higher fan speed is required to keep the air stream at the
needed level. When the fan speed increase, the fan noise will also increase.
If the cooling air stream is insufficient to stabilize the interior temperature within the operating
margin, the system will stop operating.
- Solution: Clean or replace the filters.
If a fan is worn out, it may be noisy, or stop working.
- Solution: Replace the fan(s).