GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
2 - 8 Section 2-2 - General console requirements
2-2-4 Probes environmental requirements
2-2-5 Time and manpower requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
Properly dress peripheral
cables
Do not allow cables to lie across the top of the Card Rack or hang out of the peripheral
bays.
Loop the excess length for peripheral cables inside the peripheral bays. attach the
monitor cables to the frame.
Table 2-4 Operation and storage temperatures for probes
Electronic PAMPTE
Operation: 10 to 40 ºC (50 to 104 ºF) 5 to 42.7 ºC (41 to 108,9 ºF)
Storage: -20 to 50 ºC (-4 to 122 ºF) -20 to 60 ºC (-4 to 140 ºF)
Temperatures in degrees Celsius (
o
C) conversion to degrees F: (
o
F) = (
o
C * 9/5) + 32
CAUTION
PAMPTE probes are designed for storage temperatures of -20 to +60 degrees C
(-4 to +140 degrees F).
Electronic probes are designed for storage temperatures of -20 to +50 degrees C
(-4 to +122 degrees F).
When exposed to large temperature variations, the product should be kept at room temperature
the needed time to stabilize its temperature before use.
Refer to Table 2-1 "VIVID E9 Acclimate Time" on page 2-2 to determine the needed settlement
time.
WARNINGWARNING
HAVE TWO PEOPLE AVAILABLE TO DELIVER AND UNPACK THE VIVID E9.
ATTEMPTS TO MOVE THE UNIT CONSIDERABLE DISTANCES OR ON AN INCLINE BY
ONE PERSON COULD RESULT IN INJURY OR DAMAGE OR BOTH.
Table 2-3 EMI prevention/abatement (cont’d) sheet 2 of 2
EMI RULE DETAILS