3 - 2 Section 3-2 - Setup reminders
Section 3-2
Setup reminders
3-2-1 Average setup time
3-2-2 Setup warnings
Table 3-1 Average installation time
DESCRIPTION
AVERAGE
INSTALLATION
TIME
COMMENTS
UNPACKING THE SCANNER 0.5 HOUR
INSTALL SCANNER WO/OPTIONS 4 HOURS DEPENDENT ON THE CONFIGURATION
DICOM NETWORK CONFIGURATION 2 HOURS OR MORE DEPENDENT ON THE CONFIGURATION
INSTALL INSITE / ILINK 0.5 HOUR
DANGER
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
CAUTION
IF THE UNIT IS VERY COLD OR HOT, DO NOT TURN ON ITS POWER UNTIL IT HAS HAD A
CHANCE TO ACCLIMATE TO ITS OPERATING ENVIRONMENT.
Table 3-2 VIVID E9 Acclimate Time
o
C
-40-35-30-25-20-15-10-5 0 5 10152025303540455055606570
o
F
-40 -31 -22 -13 -4 5 14 23 32 41 50 59 68 77 86 95 104 113 122 131 140 149 158
Hrs 2018161412108642000000024681012
CAUTION
TO PREVENT ELECTRICAL SHOCK, CONNECT THE UNIT TO A PROPERLY GROUNDED
POWER OUTLET. DO NOT USE A THREE TO TWO PRONG ADAPTER. THIS DEFEATS SAFETY
GROUNDING.
CAUTION
DO NOT WEAR THE ESD WRIST STRAP WHEN YOU WORK ON LIVE CIRCUITS AND MORE
THAN 30 V PEAK IS PRESENT.
CAUTION
DO NOT OPERATE THIS UNIT UNLESS ALL BOARD COVERS AND FRAME PANELS ARE
SECURELY IN PLACE. SYSTEM PERFORMANCE AND COOLING REQUIRE THIS.