GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 10 - Care & maintenance 10 - 3
Section 10-3
Maintenance task schedule
10-3-1 How often should maintenance tasks be performed?
The Care and Maintenance task schedule (provided on page 10-3) specifies how often your VIVID E9
should be serviced and outlines items requiring special attention.
NOTE: It is the customer’s responsibility to ensure the VIVID E9 care & maintenance is performed as
scheduled in order to retain its high level of safety, dependability and performance.
Your GE Service Representative has an in-depth knowledge of your VIVID E9 ultrasound scanning
system and can best provide competent, efficient service. Contact GE for coverage information and/or
price for service.
The service procedures and recommended intervals shown in the Care and Maintenance Task
Schedule assumes that you use your VIVID E9 for an average patient load (10-12 per day) and not use
it as a primary mobile unit which is transported between diagnostic facilities.
NOTE: If conditions exist which exceed typical usage and patient load, then it is recommended to
increase the periodic maintenance frequencies.
.
Table 10-1 Customer Care Schedule sheet 1 of 2
Service at Indicated Time Daily Weekly Monthly
Per
Facilities
QA
Program
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
Clean Air Filters
more frequently depending on
your environment
Clean Console
Inspect AC Mains Cable
Mobile Unit Check Weekly
Inspect Cables and Connectors
Clean Monitor and Touch Panel
Inspect Wheels, Casters, brakes and
Swivel Locks
Mobile Unit Check Daily
Check Operator Panel Movement
Mobile Unit Check Daily
Console Leakage Current Checks
also after corrective
maintenance
Peripheral Leakage Current Checks
also after corrective
maintenance
Surface Probe Leakage Current Checks
also after corrective
maintenance
Endocavity Probe Leakage Current
Checks
also after corrective
maintenance