GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 4 - General procedures and Functional checks 4 - 19
4-2-7 Recording important settings and parameters
4-2-7-1 Overview
This subsection includes descriptions for recording data from the following screens:
• Connectivity
- Dataflow (see: 4-2-7-2 "Connectivity — Dataflow" on page 4-20)
- Additional Outputs (see: 4-2-7-3 "Connectivity — Additional Outputs" on page 4-21)
- Tools (see: 4-2-7-4 "Connectivity — Tools" on page 4-22)
- Formats (see: 4-2-7-5 "Connectivity — Formats" on page 4-22
- Tcpip (see: 4-2-7-6 "Connectivity — Recording the TCP/IP settings" on page 4-23)
•System
- Settings (see: 4-2-7-7 "System — Settings" on page 4-25)
• About
- System Version (see: 4-2-7-9 "About — System Version" on page 4-26)
•Admin
- Disk Management (see: 4-2-7-13 "Admin — Disk Management" on page 4-27)
- Backup (see: 4-2-7-14 "Admin — Backup" on page 4-27)
- Restore (see: 4-2-7-15 "Admin — Restore" on page 4-28)
- Users (see: 4-2-7-16 "Admin — Users" on page 4-28)
- System Admin (see: 4-2-7-17 "Admin — System Admin" on page 4-29)
• Service
- Service screen (see: 4-2-7-18 "Service screen" on page 4-30)
NOTICE
An error, or a power loss may occur during the software loading.
Always keep a record of the settings for the VIVID E9 on paper. Verify if it is current before
starting a software loading!