8 - 282 Section 8-11 - Back End Processor (BEP) parts replacement
8-11-9 Patient I/O Interface Cable replacement
The Patient I/O Interface Cable connects the Patient IO to the BEP’s motherboard.
It is included in the BEP6.X Cable Kit - Spare Part, Part Number 5433408-80.
8-11-9-1 Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Disconnect all probes.
4.) Remove the Left Side Cover.
5.) Remove the BEP Cover.
a.) Loosen the BEP Cover’s fixing screw(s).
- On BEP6 there are two finger screws on the top of the BEP Cover.
- On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
- On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the BEP Cover away from the BEP.
c.) Set the BEP Cover aside.
6.)
8-11-9-2 Patient I/O Interface Cable removal
Follow these steps to remove the Patient I/O Interface Cable:
1.) Remove the BEP Power Board. This is required to get access to the Patient I/O Interface Cable.
2.) Unplug the two connectors (Power and USB) from the Patient I/O.
3.) Unplug the connector from CN USB11 on the BEP’s motherboard.