GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics / troubleshooting 7 - 13
Section 7-8
Motor Controller Test
This test program tests the XYZ Motor Controller. It also includes a software recovery procedure for
Lock Release.
Follow this procedure to start the Motor Controller Test:
1.) Power up the VIVID E9,.
See 4-2-1 "Power ON/Boot Up" on page 4-4 for detailed instructions.
After scanner initialization is complete, continue with the steps below;
2.) Press CONFIG on the Utility tab on the Touch Panel.
3.) Log on as ADM. A password for the user ADM may be required.
See 4-2-5 "Logging on to VIVID E9 as ‘ADM’" on page 4-15 for detailed instructions.
4.) Select System from the bottom of the menu that appears on the monitor.
5.) Select Test.
6.) Select MotorController Test
The following dialog is displayed on the screen.
Figure 7-2 Test screen