8 - 196 Section 8-8 - Main Console parts replacement
8-8-4 Bottom “nylon strip” Air Filter replacement
8-8-4-1 Overview
The bottom air filter is held in place with magnets. A tab made of Nylon (Nylon Strip) extends from the
air filter. The tab can be accessed on the right side of the VIVID E9.
8-8-4-2 Preparations
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Kneeling down on the right side of the VIVID E9 enables you to access the air filter tab.
4.) Locate the tab extending from the air filter on the bottom of the VIVID E9.
Figure 8-200 Bottom “nylon strip” Air Filter
CAUTION
Lock the VIVID E9’s wheels prior to removing/cleaning the air filter. This prevents the VIVID E9
from moving unexpectedly.