8 - 162 Section 8-6 - Top Console Parts Replacement
8-6-30 Probe Cable Hook Twin replacement
8-6-30-1 Remove the Cable Hook, Twin
1.) Unscrew the fastening screw. (It has been locked with Lock-Tite, so you may need to apply a little
extra force to unscrew the screw.)
2.) Remove the Cable Hook, Twin.
8-6-30-2 Install the Cable Hook, Twin
1.) Position the Cable Hook, Twin as shown in Figure 8-171 "Cable Hook, Twin" on page 8-162.
2.) Fix it in place by using an M5x20 screw with an M6 washer. Lock-Tite has to be used.
Torque: 50 Ncm (fingertight).
Figure 8-171 Cable Hook, Twin
SCREW, M5X20 WITH M6 WASHER