Chapter 8 - Replacement procedures 8 - 283
8-11-9-3 Patient I/O Interface Cable installation
Follow these steps to install the Patient I/O Interface Cable:
1.) Connect the two connectors (Power and USB) to the Patient I/O.
2.) Connect the other end of the cable to CN USB11 on the BEP’s motherboard.
3.) Install the BEP Power Board.
4.) Install the BEP Cover.
5.) Install the Left Side Cover.
CAUTION
BE AWARE OF CORRECT ORIENTATION OF THE PATIENT I/O CABLE CONNECTOR WHEN
CONNECTING TO CN USB11 ON THE BEP'S MOTHERBOARD.
ON SOME BEPS, THE KEYING TO PREVENT ERROR MAY BE MISSING.