GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 5 - Components and functions (theory) 5 - 79
5-10-2-1 General description (cont’d)
The mains cord has plugs in both ends. A female plug connects to the scanner and a male plug to the
wall outlet.
5-10-2-2 Temperature Control
The Main Power Supply is equipped with an internal fan with variable speed for temperature control.
Both the temperature of the air entering the power supply and leaving the power supply are measured.
The fan is controlled by software.
5-10-2-3 Input
Mains Power, 110 VAC or 230 VAC, 50/60 Hz
Figure 5-57 Main Power Supply
Figure 5-58 L1 - Mains Inlet connector
Main Power Supply