EasyManua.ls Logo

GE Vivid E9

GE Vivid E9
802 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 5 - Components and functions (theory) 5 - 81
5-10-2-6 Fuses
Ceramic body fuses inside the power supply. (Only to be replaced by the Main Power Supply
manufacturer.)
Fuses data:
15 A Fast Acting type
5-10-2-7 Current limiter, over-voltage protection and temperature watch-dog
A current limiter will switch off the power if any of the outputs are overloaded.
Over-voltage Protection is provided for these voltages:
- + 3.3 V
-+ 5 Vd
-+/- 5 Va
- +/- 15 Va
Voltage will be turned off if the temperature grows too high (temperature watch-dog).
P5
+48 VDC
BACK END PROCESSOR AND MOTOR
POWER
- The ‘48V_OK’ indicates that the 48 VDC
measured on the BEP, is OK.
- The ‘+5Vstb’ is always ON, and is connected to
the BEP’s motherboard and to the light in the ON/
OFF button on the Operator Panel.
P6
See: Table 5-23 "USB connector" on page 5-80. BACK END PROCESSOR
P7
GND STUD GROUND STUD
Table 5-24 Power outputs (cont’d) sheet 2 of 2
CONNECTOR DESCRIPTION COMMENTS

Table of Contents

Other manuals for GE Vivid E9

Related product manuals