5 - 90 Section 5-11 - Input and Output (I/O) modules
5-11-2-3 Input DC Voltages
+5 VDC
+12 VDC
Pinout for the DC input on the Patient I/O module
5-11-2-4 Patient I/O - inputs
ECG / Respiration
Phono (from a phono heart microphone)
AUX - Analog Input (Pulse/Pressure)
Pinout for the AUX connector
The pinout for the AUX connector is described in the table below:
The AUX is default a 1 Vpp (Volt peak-to-peak) input with a max frequency of 300 Hz. It has a
programmable high gain mode with a maximum input signal of 300 mVpp (millivolt peak-to-peak).
The inputs are differential. For a single ended sensor signal, the pin 1 (input -) should be connected to
the GND of the sensor.
Table 5-25 DC input pinout on the Patient I/O Module
CONNECTOR SIGNAL NAME
DC INPUT
Pin 1: +12VDC
•Pin 2: GND
•Pin 3: GND
•Pin 4: +5VDC
Table 5-26 Pinout for the AUX Connectors
CONNECTOR SIGNAL NAME
AUX
Pin 1: Input -
Pin 2: Input +
•Pin 3: Gnd
Pin 4: Nasal Sensor 1
Pin 5: Nasal Sensor 2
4321
1
2
34
5