GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics / troubleshooting 7 - 23
7-10-2-3 Different Location
Move the scanner to another location and verify if the noise changes or disappear. This may help you
to locate an external noise source.
Try to move the scanner to:
another location inside the room
another room
another floor
7-10-2-4 Disconnect External Cables
1.) Disconnect all external cables (network, all unused probes, ECG leads and verify if the noise
disappears.