Chapter 8 - Replacement procedures 8 - 15
8-4-9-3 Initiate software reload from the Recovery Console (cont’d)
If, accepted, the screen below is displayed.
6.) Press the Enter key on the keyboard. The software reload starts.
NOTE: Time to complete the software reload is typically 15 to 25 minutes.
- When the System Software reload is done, the VIVID E9 reboots.
- After the reboot, the VIVID E9 starts to load the Application Software. This is done without any
interactions. The TCP/IP settings, Computer Name etc. are restored.
- The VIVID E9 shuts down.
Figure 8-4 Reload System Software