8 - 68 Section 8-5 - Replacing Covers and Bumpers
8-5-16-2 Covers under XY / Frogleg motors removal procedure (cont’d)
2.) Under the XY / Frogleg mechanism, identify the 4 separate covers. There are two types of covers,
one type has two screws (1) and the other type has one screw (2).
3.) Remove the screws from the cover(s) to replace.
4.) Pull down and slide the cover away from the XY mechanism. Be sure to flex the plastic slightly so
the plastic clears the XY.
5.) Disconnect the cable.
Figure 8-69 XY / Frogleg mechanism covers, right side frog leg, from underneath
Figure 8-70 Pull down and slide XY cover out
PLASTIC MUST
CLEAR FOR THE
COVER TO
RELEASE AND
SLIDE OUT