Chapter 8 - Replacement procedures 8 - 213
8-9-2-4 Front Casters installation procedure
1.) Position the Caster so it align with the fastening screw.
2.) Install the three fixing screws (M8 x 20 mm), Torque: 20.5 Nm.
3.) Remove the Wooden Wedge.
4.) Roll the VIVID E9 off the Bevel Edged Board.
5.) To replace the other Front Caster, repeat all the steps from 8-9-2-3 "Front Casters removal
procedure" on page 8-212, but now using the Bevel Edged board and the Wooden Wedge on the
other side of the VIVID E9.
6.) Install the Foot Rest Cover.