Chapter 8 - Replacement procedures 8 - 245
8-11-1-2 Preparations
When preparing for the replacement, you must perform the following steps:
1.) Record the VIVID E9’s TCPIP settings and installed Option strings.
2.) Export the Patient Archive images to a server or to external media.
3.) Back up the Report Archive, User-Defined Configuration (Customer Presets), and Service
settings.
4.) If possible, wipe the HDD partitions as described here: Reload the system software from DVD.
Select A: to overwrite all content on HDD.
5.) Power down the VIVID E9.
6.) Disconnect the Mains Power Cable from the wall outlet.
7.) Disconnect all probes.
8.) Remove the Side Covers.
9.) Remove the Top Cover.
10.)Remove the B/W Printer.
- Disconnect the Power Cable from the rear of the printer.
- Disconnect the signal cable from the I/O panel.
- Loosen the printer bracket wing nut.
- Slide the printer (with signal cable) forward, out of the VIVID E9.
11.)For easier access in the next steps, remove the VIVID E9’s Front Cover.