GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
8 - 306 Section 8-14 - Peripherals replacement
8-14-2-2 Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Disconnect all probes.
4.) Remove the Left Side Cover.
5.) Remove the BEP Cover.
a.) Loosen the BEP Cover’s fixing screw(s).
- On BEP6 there are two finger screws on the top of the BEP Cover.
- On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
- On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the BEP Cover away from the BEP.
c.) Set the BEP Cover aside.
6.) For easier access to the DVR Board, remove the Hard Disk Drive.
Next (select what applies):
• 8-14-2-3 "BEP6 - Digital Video Recorder removal procedure" on page 8-307.
• 8-14-2-4 "BEP5 - Digital Video Recorder removal procedure" on page 8-308.