Chapter 10 - Care & maintenance 10 - 19
10-5-6-6 Disinfecting probes (cont’d)
NOTE: For the 13L probe, please refer to the User’s Manual for Intraoperative Probes, Direction
Number: KX192055.
3.) Rinse the part of the probe which was in contact with the germicide according to the germicide
manufacturer's instructions.
4.) Wipe dry with a soft towel or air dry the probe.
Figure 10-2 Probe immersion levels
A. 3V-D, 4V-D, M5S-D, 6S-, 12S-D
B. 4C-D
C. 9L-D, 11L-D, ML6-15-D
D. 2D (P2D)
E. 6D (P6D)
1. Fluid level
2. Contact face with patient environment
WARNINGWARNING
CREUTZFELD-JAKOB DISEASE
NEUROLOGICAL USE ON PATIENTS WITH THIS DISEASE MUST BE AVOIDED. IF A
PROBE BECOMES CONTAMINATED, THERE IS NO ADEQUATE DISINFECTING
MEANS.