Error
code
Error Note
A04 Chamber empties
slowly.
Level sensor in the chamber (-B21) is not activated after 2
tries at emptying.
Possible action:
a. Check the function of the drainage pump.
b. Check the function of the level sensor.
c. Check that water can be pumped out of the drainage
system.
A05 Water leakage in the
dryer.
The sensor in the drying box (-B09) is activated.
Possible action:
a. Check the check valves in the drying channels.
b. Check the function of the sensor.
A06 Unlocked door. The door locking switch (-S01) is not activated when the
machine is running a process.
Possible action:
a. Check that nothing is trapped in the door.
b. Check that the door locking switch (-S01) is activated
on door locking.
A14 Low flow of detergent 1. Flowmeter (-B17) registers low flow (2 ml/5 s) from the dosing
pump (-M03).
Possible action:
a. Check the functioning of the dosing pump.
b. Check that there is detergent in the container.
c. Check that there is no air in the dosing system.
d. Check that the hoses are not blocked.
A15 Low flow of detergent 2. Flowmeter (-B18) registers low flow (2 ml/5 s) from the dosing
pump (-M04).
Possible action:
a. Check the functioning of the dosing pump.
b. Check that there is detergent in the container.
c. Check that there is no air in the dosing system.
d. Check that the hoses are not blocked.
A16 Low flow of detergent 3. Flowmeter (-B19) registers low flow (2 ml/5 s) from the dosing
pump (-M05).
Possible action:
a. Check the functioning of the dosing pump.
b. Check that there is detergent in the container.
c. Check that there is no air in the dosing system.
d. Check that the hoses are not blocked.
4 ALARM MANAGEMENT
TECHNICAL MANUAL GETINGE WD15 CLARO 57 (128)
<Doc_TEC><Doc_6001396902><Rev.I><Lang_en>